|
Selective head
cooling involves cooling the head while providing radiant warmth to the
remainder of the body and allows the brain to be cooler than the rest of
the body. This is accomplished by maintaining the radiant warmer output
at 100% and adjusting the cap temperature to achieve the target rectal
temperature of 34.5°C ± 0.5° for the 72 hour cooling period.
Standard
Criteria for Head Cooling:
The infant should
be at least 36 weeks GA, ≥1800 grams, and less than 6 hours old with:
- History
compatible with HIE: Apgars <5 at 10 min, pH <7, resuscitation at 10
min, base deficit ≥16 (Note: Should meet ONE of these)
- Physical exam
compatible with HIE: lethargy, hypotonia, seizures, coma, weak suck
(Note: Should meet ONE of these as well)
- An EEG will
be done at time of admission
Starting Head
Cooling:
Keep the radiant
warmer off while the infant is being assessed for cooling and setting up
the Olympic Cool-cap system.
Disposable
items from service center:
- #86126 small
cap set (HC <32cm, weight 1.8-2.5kg)
- #86127 medium
cap set (HC 32-37cm, weight 2.5-4kg)
- #86128 large
cap set (HC >37cm, weight >4kg)
- #86130 box of
temperature probes (rectal, scalp, and skin) and sensor covers
- EEG needle
electrodes- non chargeable
- 1000ml
sterile water inhalation bag (same bag RT uses for vents)
Temperature
probes:
-
Rectal temperature probe –
Mark rectal temp probe at 5 cm, lubricate, and insert tube to mark,
secure with tape. Check placement of rectal probe prn or if rectal
temp changes. If probe comes out, clean, lubricate, and replace.
-
Scalp- place temperature probe plug
into sensor to proceed (must be placed into sensor even though not
placed on baby, no longer using scalp temp probe over fontanel)
Place cooling plastic cap over head, open to the face, not past the
hairline, centered with the cutouts over the ears, then place the
blue and silver caps over the infant’s head with the silver cap not
to extend past the chin, make sure the strap is not stretched too
tight under the chin.
-
Radiant warmer and cooling machine
skin temperature probes – need to be placed over the liver area with
reflector probe covers.
General baby
care: Keep the bed flat with
the infant horizontal, no diaper on infant to allow maximum body
exposure to the radiant warmer (can use chux, urine bag, or catheter),
do not swaddle in snuggly or blanket rolls, do not blow or circulate air
over infant and no enteral feeds during the cooling and rewarming
processes of treatment. Keep the reflective Plexiglas shield over the
infant’s head at the level of the chin/neck, resting evenly and directly
on the mattress. It helps to keep the infant close to the head of the
bed so the shield covers the cooling tubing. Do not block the front or
back openings of the heat shield.
Initial cool
down: Record baseline vitals
and temperatures on the Temperature Flow Sheet (place in MR in progress
notes when cooling completed). With the radiant warmer off, turn on the
Cool-cap system with initial cap temperatures of:
-
Infant’s < 2.5kg: 12°C to 15°C
-
Infant’s 2.5kg – 4kg: 10°C to 12°C
-
Infant’s > 4kg: 8°C to 10°C
The cap
temperature cannot be set lower than 8°C.
The rectal
temperature is controlled by changes to the cap temperature with a
temperature goal of 34.5°C rectally. Changes are made to the cap temp
in 0.5° increments or less. Wait up to 45 minutes for the infant’s
rectal temperature to respond. It is important to notice trends.
To adjust the
cap temperature:
-
Touch Set Cap. The set cap
water temperature window appears.
-
Touch ˆ to increase or ˇ to
decrease the water temperature. Each button touch will increase or
decrease the desired cap water temp by 0.1°C. Normal adjustments
should be made in increments of 0.5° or less and, preferably, in
increments of 0.1°C. A change of 0.1°C will cause an approximate
0.1°C change in rectal temperature.
-
Touch OK to save the new
temperature.
When the rectal
temperature reaches 35.5°C, turn on the radiant warmer in the SERVO
mode. The initial radiant warmer servo setting should be 37°C. If
the rectal temp is already 35.5°C or lower on admission, the radiant
warmer may be turned on to prevent overcompensation. Once the skin
temperature stabilizes, regularly adjust the servo setting to
approximately 0.5°C above the skin temperature This maintains the
radiant warmer output at approximately 100% as the infant’s temperature
changes. Never set the radiant warmer control servo temp > 37.5°C.
to prevent overheating the infant. Skin temp can be adjusted for
small infants (<2.5 kg): 37 – 37.5°C, average infants (2.5-4kg): 36-37 °
C and large infants (>4kg): 35 - 36°C. While it is acceptable for the
radiant warmer output to occasionally drop to 70% to 80% for short
periods of time, the radiant warmer should be adjusted to remain at 100%
output most of the time.
DO NOT USE THE
RADIANT WARMER IN THE MANUAL MODE
as it will make the infant’s temperature less stable.

Controlling the
radiant warmer during cooling maintenance:
Pausing cooling
treatment for scalp checks:
-
During the
cooling treatment, check the integrity of the scalp at least every
12 hours. The system will prompt for these scalp checks every 12
hours. It is normal to see an impression of the water cap on the
scalp. Check for redness, bruising, and other discoloration. Pause
cooling treatment during scalp checks. Whenever possible, limit
periods of cap removal to 30 minutes or less. To avoid disruption
of stable cap temperature settings while removing the cap for short
term checks, always Touch pause before removing the caps when
cooling treatment is in process.
To pause the
cooling treatment:
-
Touch PAUSE
-
If baby is remaining in bed with
cap off and paused for EEG or ultrasound be sure to turn down
radiant warmer, try to maintain rectal temperature in target range
-
Remove the heat shield
-
Remove the complete cap set from
the infant’s head, then perform the medical procedures and scalp
checks. (If the infant needs to leave the floor, turn off the
radiant warmer, unplug the temperature probes from the temperature
sensor module, and leave the caps and module with the cooling
machine. When the infant returns, plug the temperature sensors in,
turn the radiant warmer back on and continue with the next step)
-
Place the complete cap set back on
the infant’s head
-
Place the heat shield
-
Touch RESUME to continue the
treatment, if resume is not pressed within 10 minutes to continue
cooling treatment, the system will alarm. If additional time is
required, touch OK.
Rewarm
Procedure:
During the rewarm
procedure, the goal is to raise the infant’s rectal temperature 0.5°C
per hour in 0.2°C and 0.3°C increments every 30 minutes. This procedure
may take up to 4 hours if starting at a rectal temperature of 34.5°C -
35°C. Larger infant’s may require less heat to rewarm or may rewarm
faster than 0.5°C per hour, despite the radiant warmer setting. Closely
monitor these infants. Regularly check the infant’s rectal temperature
between the 30 minute servo adjustments.
To rewarm the
infant:
-
A message box displays when the
72-hour cooling treatment is complete. TOUCH OK to access
the rewarm wizard. Touching DONE will confirm the task as
complete and display the next task.
-
Remove the heat shield and set
aside for cleaning
-
Remove the 3 caps: insulating,
water retainer, and water cap. Open the Velcro chin strap, and then
lift all 3 caps, with the tubing, off the infant’s head. Discard
the blue water cap retainer and silver insulating cap. Keep the
water cap and cap connector tubes connected to the system; these are
required for the shutdown procedure.
-
Leave the temperature sensors in
place; the system will continue to monitor the infant’s temperatures
during the 4 hour rewarm period.
-
Adjust the radiant warmer servo
temperature to approximately 0.3°C above the rectal temperature (or
35°C, lowest Giraffe bed will go, if increasing the temp more than
0.3°C, adjust or skip changing servo temp at the first 30 minute
prompt)
-
Confirm the tasks as complete. The
system stops circulating and cooling water. TOUCH REWARM to
begin the rewarm period.
-
During the 4 hour rewarm period,
the system prompts for adjustment of the radiant warmer servo
setting 0.2°-0.3°C every 30 minutes. The system will automatically
stop when the 4 hour rewarm period is complete.
-
The Cool-cap system uses a 4 hour
rewarm period, to shorten this period, touch PATIENT, and
then touch SHUTDOWN to stop the rewarm period.
-
After the 4 hour rewarm period is
complete, the system can be shutdown. TOUCH OK to access the
shutdown wizard. Follow the prompts on the screen
Ending the
cooling treatment early:
If a medical
treatment requires that cooling treatment be stopped prematurely,
rewarming is recommended.
- TOUCH
PATIENT, then touch
SHUTDOWN
- Touch
REWARM and follow the on-screen instructions to rewarm the
infant.
- Touch
SHUTDOWN and follow the on-screen instructions to bypass
rewarming and shut down the system.
- Touch
POWER DOWN to bypass both the rewarm and shutdown wizards and
return to the start screen.
Off-Line
Rewarming:
If it is necessary
to rewarm the infant without access to a Cool-cap system, use the
following procedure to rewarm the infant to a rectal temperature of
36.5°C.
- Remove the
heat shield
- Remove the 3
caps, open the Velcro chin strap, and then lift all 3 caps with the
tubing, off the infant’s head.
- Discard the
blue water cap retainer and silver insulating cap, keep the water
cap and cap connector tubes connected to the system for the shutdown
procedure.
- Remove the
rectal and skin temperature sensors, replace servo temp probe over
liver area
- Insert
Philips rectal probe and connect to Philips monitor
- Adjust the
radiant warmer servo temperature approximately 0.3°C above the
infant’s rectal temperature, or 35°C
- After 30
minutes from the initial servo temperature change, increase the
radiant warmer servo temperature by 0.2°C, adjust if first increase
> 0.3°C, record the time, rectal temp, and radiant warmer servo
setting.
- After 30
minutes, increase the radiant warmer servo temperature by 0.3°C;
record the time, rectal temperature, and radiant warmer servo
setting.
- Continue to
repeat these 2 steps until the infant’s rectal temperature reaches
36.5°C. Do not set the radiant warmer servo temperature > 37.5°C
- If rewarm was
in progress, alternate increasing the radiant warmer servo
temperature in 0.2°C and 0.3°C increments.
PRECAUTIONS AND WARNINGS
Cooling:
-
Hypothermia can inhibit the
metabolism and clearance of many anticonvulsants (including
phenobarbitone, phenytoin, Lidocaine, and the benzodiazepines)
potentially prolonging their half-lives. Monitor blood levels.
-
Hypothermia inhibits antimicrobial
activity, it is important to take appropriate cultures and initiate
antibiotic treatment even in infants with apparent HIE since the
diagnosis can occasionally be confounded by peripartum sepsis.
-
Hypothermia has anticoagulant
effects.
-
At very low temperatures (30°-32°C)
hypothermia may increase PPHN
-
If an infant is markedly overcooled
(~32°C) they may reduce their peripheral circulation in an attempt
to conserve heat. This will reduce the effectiveness of the radiant
warmer, thus increasing the problem. Should this occur, remove the
cap and carefully rewarm the infant <0.5°C per hour until the infant
is within the target range. Then resume cooling using the previous
or a slightly higher cap setting.
-
Hypothermia is known to increase
oxygen consumption; therefore some infants may show a small increase
in their oxygen requirement.
-
Hypothermia may produce
prolongation of the QT interval in infants with bradycardia but
should normalize after cooling. Be cautious with other treatments
that may further prolong the QT interval (drugs and electrolytes).
-
Full term infants with a birth
weight < 2.5kg and/or infants with cardiac compromise requiring
inotrope support may require a higher cap temperature during the
initial cool down and should be closely monitored. Make temperature
adjustments early to avoid over- shooting.
-
Treatment of infants < 1.8kg is
contraindicated.
Changes in
Temperature:
-
Seizures have unpredictable
effects. They usually cause an increase in temperature, but have
also been known to decrease temperature.
-
Anticonvulsants are likely to lower
the rectal temperature
-
Core temperature will increase in
infants who are upset, jittery, or shivering.
-
Core temperature will decrease in
infants who are quiet and sleeping.
-
Correct all blood gases and pH
values for core temperature.
-
If the rectal temperature is too
high and cap temperature will not reach 8°C, it may help to lower
the room temperature by 1°C
Radiant Warmer:
-
DO NOT USE THE RADIANT WARMER IN
THE MANUAL MODE
-
Do not set the radiant warmer servo
temperature above 37.5°C to prevent overheating infant.
-
The infant’s rectal temperature
takes approximately 45 minutes to stabilize after the cap
temperature has been adjusted
Rewarming:
-
During rewarm, regularly check the
infant’s rectal temperature between the 30-minute servo adjustments.
-
Active rewarming should not exceed
0.5°C per hour.
-
Larger infants may require less
heat to rewarm or may rewarm faster than 0.5°C per hour
-
Small, full term infants weighing <
2.5kg may require special attention during rewarming. Watch
carefully for cardiovascular instability.
-
Ensure that the infant does not
rewarm at more than 0.5°C per hour when re-exposed to the full force
of the radiant warmer. Rewarming too quickly can cause vasodilation
and hypotension in an unstable infant.
|